Standards | Class 1: IEC 61672-1:2013 |
Number of Profiles | 3 |
Weighting Filters | A, B, C, Z |
Time constants | Slow, Fast, Impulse |
RMS Detector | Digital True RMS detector with Peak detection, resolution 0.1 dB |
Microphone | ACO 7052E, 35 mV/Pa, prepolarised 1/2” condenser microphone |
Calibration | Automatic calibration start at 114dB/1kHz |
Preamplifier | SV 12L IEPE type |
Linear Operating Range | 25 dBA RMS ÷ 140 dBA Peak (in accordance to IEC 61672) |
Total Dynamic Measurement Range | 15 dBA RMS ÷ 140 dBA Peak (typical from noise floor to the maximum level) |
Internal Noise Level | less than 15 dBA RMS |
Dynamic Range | >110 dB |
Frequency Range | 10 Hz ÷ 20 kHz with ACO 7052E microphone or
10 Hz ÷ 40 kHz with optional microphone, e.g MK 202E |
Meter Mode Results | SPL, Leq, SEL, Lden, Ltm3, Ltm5, LMax, LMin, LPeak, “sliding Leq” up to 60 minutes Simultaneous measurement in three profiles with independent set of filters and detectors |
Statistics | Ln (L1-L99), complete histogram in meter mode, linear of with detector (exponential)
Ln (L1-L99) statistics in 1/1 or 1/3 octaves (if option is active) |
Data Logger | logging of summary results with step from 1s
logging of time history with step from 2ms |
Audio Recording (option) | Audio events recording during measurement, trigger or continuous mode, 12 / 24 / 48 kHz sampling rate, 16 or 24 bit, wav format |
Wave Recording (option) | Time domain signal recording during measurement 12 / 24 / 48 kHz sampling rate, 16 or 24 bit, wav format |
Sound Analyser |
|
1/1 Octave Analysis (included) | Real-time analysis meeting Class 1 requirements of IEC 61260, center frequencies from 31.5 Hz to 16 kHz (audio band) with selectable weighting filter and detector or
Real-time analysis meeting Class 1 requirements of IEC 61260, center frequencies from 1 Hz to 31.5 kHz (ultrasound band) with linear detector |
1/3 Octave Analysis (option) | Real-time analysis meeting Class 1 requirements of IEC 61260, center frequencies from 20 Hz to 20 kHz (audio band) with selectable weighting filter and detector or
Real-time analysis meeting Class 1 requirements of IEC 61260, center frequencies from 0.8 Hz to 40 kHz (ultrasound band) with linear detector |
FFT Analysis (option) | 400 or 800 or 1600 lines in selectable band from 78 Hz to 20 kHz with selectable weighting filter, selectable averaging: linear or exponential, and selectable window or
400 or 800 or 1600 lines in ultrasound band from 25 Hz to 40 kHz with selectable averaging: linear or exponential, and selectable window |
RT 60 (option) | Reverberation time analysis in 1/1 octaves, selectable method: decay or impulse, or
Reverberation time analysis in 1/3 octaves, selectable method: decay or impulse |
Vibration Level Meter and Analyser |
|
Standards | ISO 10816-1 |
Number of Profiles | 3 |
Weighting Filters | Acceleration: HP1, HP3, HP10; Velocity: Vel1, Vel3, Vel10, Displacement: Dil1, Dil3, Dil10
Machine filter: VelMF (ISO 10816-1) |
RMS Detector | Digital True RMS detector with Peak detection, resolution 0.1 dB Time constants: from 100 ms to 10 s |
Accelerometer (option) | SV 80 IEPE, single axis, sensitivity 10 mV/(ms-2) ~ 100 mV/g |
Measurement Range | 0.01 ms-2 RMS ÷ 500 ms-2 Peak (with SV 80) |
Frequency Range | 1 Hz ÷ 5 kHz (with SV 80 and HP or HP1 filter) |
Vibration Analyser |
|
1/1 Octave Analysis (included) | Real-time analysis meeting Class 1 requirements of IEC 61260, center frequencies from 1 Hz to 16 kHz (full band) with selectable weighting filter and detector or
Real-time analysis meeting Class 1 requirements of IEC 61260, center frequencies from 1 Hz to 31.5 kHz (ultra band) with linear detector |
1/3 Octave Analysis (option) | Real-time analysis meeting Class 1 requirements of IEC 61260, center frequencies from 0.8 Hz to 20 kHz (full band) with selectable weighting filter and detector or
Real-time analysis meeting Class 1 requirements of IEC 61260, center frequencies from 0.8 Hz to 40 kHz (ultra band) with linear detector |
FFT Analysis (option) | 400 or 800 or 1600 lines in selectable band from 78 Hz to 40 kHz with HP weighting filter, selectable averaging: linear or exponential, and selectable window |
RPM (option) | Rotation measurement option without Laser Tachometer (option), measurement of RPM, RPS, RMAX, RMIN |
General information |
|
Input | TNC (IEPE type) |
Memory | MicroSD card 8 GB (removable & upgradeable) |
Display | Colour 320 x 240 pixels, OLED type |
Keyboard | 9 push buttons |
Communication Interfaces | USB 2.0 client |
Bluetooth® (option) | |
RS 232 (with SV 55 option) | |
External I/O – AC output (1 V Peak) or Digital Input/Output (Trigger – Pulse) | |
Power supply | Four AA alkaline or rechargeable NiMH batteries (not included) |
USB interface 500 mA HUB | |
External power supply 6 V/500 mA DC ÷ 15 V/250 mA DC | |
SA 17A external battery pack (option) | |
Environmental Conditions | Temperature from -10 oC to 50 oC |
Humidity up to 90% RH, non-condensed | |
Dimensions | 340 mm x 79 x 39 mm (including microphone and preamplifier) |
Weight | Approx. 0.6 kg with batteries |